Coronavirus disease 2019 (COVID-19) is a life-threatening infection caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2).1 Diabetes mellitus is one of the most frequent comorbidities, related to hospitalization due to SARS-CoV-2 infection, as well as a risk factor for disease severity, poor disease outcome, increased morbidity and mortality rate.2–6 SARS-CoV-2 enters human cells by binding […]

12 mins

The Role of Remote Glucose Management Using Real-time Continuous Glucose Monitoring Systems in ICU-hospitalized Patients with COVID-19

Georgios Logothetis, Konstantinos Avramidis, Evanthia Konstantaki, Vasiliki Matziou, John Doupis
Share
Facebook
X (formerly Twitter)
LinkedIn
Via Email
Mark CompleteCompleted
BookmarkBookmarked
Copy LinkLink Copied
Download as PDF
Published Online: Apr 28th 2025 touchREVIEWS in Endocrinology. 2025;21(1):Online ahead of journal publication DOI: 10.17925/EE.2025.21.1.7
Select a Section…
1

Abstract

Overview

The recent coronavirus disease 2019 (COVID-19) pandemic has induced many challenges in the clinical environment worldwide. In a bid to reduce the exposure of healthcare providers to severe acute respiratory syndrome coronavirus 2 and the utilization of personal protective equipment (PPE), while maintaining optimal patient care, in April 2020, the US Food and Drug Administration issued a new policy, allowing the use of continuous glucose monitoring (CGM) systems in the intensive care unit (ICU) setting. This article aimed to explore the role of real-time continuous glucose monitoring systems in patients in the ICU with COVID-19. The hybrid protocols integrating real-time CGM and point of care seem to be a feasible and safe alternative for the glycaemic management of critically ill patients with COVID-19, including the reduction of healthcare providers’ exposure and the preservation of PPE, whilst achieving and maintaining optimal glycaemic control.

Keywords
2

Article

Coronavirus disease 2019 (COVID-19) is a life-threatening infection caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2).1 Diabetes mellitus is one of the most frequent comorbidities, related to hospitalization due to SARS-CoV-2 infection, as well as a risk factor for disease severity, poor disease outcome, increased morbidity and mortality rate.2–6

SARS-CoV-2 enters human cells by binding to the angiotensin-converting enzyme receptor 2 (ACE2). The increase in blood glucose levels modifies the ACE2 receptor in a way that makes it more susceptible to viral binding.7 Hyperglycaemia itself triggers the production of inflammatory mediators (such as cytokines), which can further increase the expression or activity of ACE2 on cell surfaces, thereby making it easier for SARS-CoV-2 to attach and enter cells.8

Acute hyperglycaemia in patients with diabetes and COVID-19 infection may increase hospitalization duration and related morbidity, whereas effective glycaemic control in critically ill patients may positively affect the clinical outcomes of the disease.9,10 Furthermore, the effective management of acute hyperglycaemia in patients with COVID-19 seems to reduce viral load during the infection, although the pathogenetic mechanisms remain unknown.11 Adequate glucose level management plays an important role, especially in critically ill intensive care unit (ICU) patients, who receive continuous insulin infusion (CII) for hyperglycaemic crises, such as diabetic ketoacidosis (DKA) and hyperosmotic hyperglycaemic coma.12

CII is considered the optimal method for the management of hyperglycaemia in the ICU, as it ensures rapid onset and short action duration of the infused insulin, optimizing the glucose management of the patient in an environment with high glucose variability.13,14 CII procedure prerequisites accurate and hourly glucose measurements.15 Blood glucose assessment in hospitalized patients is usually performed by whole blood lab analysis or by point-of-care (POC) glucose measurements in arterial, venous or capillary blood samples.16 However, these measurements may delay the detection of abnormal glycaemic patterns and, consequently, affect the optimal insulin infusion rate.17 Hourly glucose measurements may also cause discomfort, blood loss and sleep disturbances.18 In addition, hourly POC measurements in patients with COVID-19 increase the workload and the risk of virus transmission to the healthcare personnel, also increasing the requirements for personal protective equipment (PPE).19 To address the new COVID-19 crisis challenges in clinical practice, the US Food and Drug Administration (FDA) has recently issued a new policy for glucose monitoring in hospitalized patients, recommending the use of non-invasive and remote monitoring devices, legitimizing the use of continuous glucose monitoring (CGM) in the hospital environment.20–22

Large randomized clinical trials and observational studies have already demonstrated the glycaemic benefits of CGM systems in both outpatient and hospitalized patients (Table 1).23–32 The use of a real-time CGM (rtCGM) system may provide real-time monitoring of the interstitial fluid glucose levels, transmitting the measurements approximately every 5 minutes. Furthermore, CGM systems provide alarms for out-of-range glucose values, enabling the timely detection of possible unrecognized hyper- or hypoglycaemic episodes, thereby reducing the need for frequent POC testing.33

Table 1: Characteristics of studies included in the systematic review23–28

Author, year

Study population

Type of CGM

Reference sample

Objective

Results

Chow et al., 202128

30 ICU patients with COVID-19

Dexcom G6 (Dexcom, San Diego, CA, USA)

Arterial POC (Nova Statstrip®Nova Biomedical, Waltham, MA, USA)

Evaluation of the accuracy and clinical benefits of rtCGM systems in the glycaemic management of patients with diabetes and severe COVID-19 infection

A 14% reduction in the average glucose sensor value during management with rtCGM, accompanied by a 50% reduction in POC measurements in 50% of patients. Additionally, 64% of nurses reported improved care for patients with diabetes, and 49% reported a reduction in the use of invasive procedures

Davis et al., 202123

9 ICU patients with COVID-19

Dexcom G6

Arterial or capillary POC (Nova Statstrip)

Evaluation of a hybrid
rtCGM-POC protocol integrated with an automated
decision-making algorithm for CII

75.7% of the sensor values fell within 20% of the reference POC values, with a 63% reduction in POC measurements. The average TIR was calculated at 71.4%. Mechanical factors affected sensor accuracy in four patients

Sadhu et al., 202024

11 ICU patients with COVID-19

Medtronic Guardian Connect (Medtronic, Northridge, CA, USA) (n=6) and Dexcom G6 (n=5)

Arterial, venous or capillary POC (Roche Accu-Chek® Inform II; Roche, Basel, Switzerland)

Evaluation of the applicability and accuracy of an rtCGM system with intermittent POC checks in the glycaemic management of critically ill patients

Both rtCGM systems were easily implemented, showing acceptable accuracy (Medtronic: MARD=13.1% with 100% of values in zones A and B of the CEG; Dexcom G6: MARD=11.1% with 98% in the corresponding zones), along with good nursing acceptance

Agarwal et al., 202125

11 ICU patients with COVID-19

Dexcom G6

POC

Evaluation of the accuracy and performance of rtCGM systems

MARD was calculated at 12.58%, median ARD at 6.3%, while the use of the rtCGM system resulted in approximately a 60% reduction in POC glucose measurements in patients receiving CII

Faulds et al., 202126

19 ICU patients with COVID-19

Dexcom G6

Arterial or capillary POC (Nova Statstrip)

Evaluation of the safety and feasibility of a hybrid model combining rtCGM and POC glucose measurements

A 71% reduction in the number of POC measurements was observed. TIR (for glucose values between 70 and
180 mg/dL) on the first day of use was 63%, while from the second day to the seventh day, it increased to 72%. Simultaneously, there was a reduction in TBR (<70 mg/dL) from 1.5% on the first day to 0.16% on the corresponding days

Longo et al., 202227

10 ICU patients with COVID-19 and 18 general ward patients*

Dexcom G6

POC (Roche Accu-Chek Inform II) and whole blood analysis from the laboratory (Beckman DxC 800; Beckman Coulter, Brea, CA , USA)

Evaluation of the accuracy, safety and feasibility of rtCGM systems

MARD in critically ill patients was calculated at 12.1% when the sensor glucose value was compared with both POC and laboratory values, 10.4% when compared only with the laboratory, and 12.7% when compared only with the POC

*In the analysis of the review, data were used only from patients in the ICU.

CGM = continuous glucose monitoring; CII = continuous insulin infusion; COVID-19 = coronavirus disease 2019; ICU = intensive care unit; MARD = mean absolute relative difference; POC = point of care; rtCGM = real-time continuous glucose monitoring; TBR = time below range; TIR = time in range.

In this narrative review article, we describe the clinical benefits and the accuracy of real-time continuous glucose monitoring systems in the glycaemic management of critically ill patients with COVID-19 infection, hospitalized in ICU settings.

Accuracy

Four studies evaluated the mean absolute relative difference (MARD) of Dexcom G6, while only one study assessed the MARD of Medtronic Guardian Connect.24–27 MARD analysis is commonly employed to assess the accuracy of CGM systems, although it is occasionally applied to blood glucose monitoring (BGM) systems, especially when comparing the two types of systems. MARD is determined by calculating the average of the absolute relative differences between the measurement results from the CGM/BGM system and the corresponding reference method.34 The overall mean MARD of both rtCGM systems was reported to be 12.47%, with Dexcom G6 performing slightly better than Medtronic Guardian Connect (Mean MARD-Dexcom G6: 12.32%; MARD-Medtronic Guardian Connect: 13.1%). The MARD value was even lower for both systems for POC glucose values greater than 180 mg/dL, compared with the overall MARD for all glucose ranges.24,26 However, available data were limited for precisely assessing the MARD for POC measurements <70 mg/dL.24,26,27 Moreover, it seemed that the first day of the use of the sensor tended to provide less accurate measurements, a fact that was also observed by Faulds et al., where the overall MARD decreased from 13.9 ± 7.8% on the first day to 13.5 ± 8.1% on the second to seventh day of sensor use.26,27 Additionally, Faulds et al. evaluated the overall mean absolute difference, which was reported as 25 ± 16.1 and 23.2 ± 15.5% for the same time periods, respectively.26

In another study, MARD of Dexcom G6 was reported to be 12.58%, while the median MARD was 6.3%.25 Furthermore, Sadhu et al., in a study investigating both Dexcom G6 and Medtronic Guardian Connect, found that Dexcom G6 demonstrated better performance in terms of accuracy (MARD 11.1%, concordance correlation coefficient 0.89 and Pearson correlation coefficient 0.9) compared with Medtronic Guardian Connect (MARD 13.1%, concordance correlation coefficient 0.79 and Pearson correlation coefficient 0.84).24

The error grid (EG) analysis, a method used to evaluate the clinical accuracy of rtCGM systems, was assessed in three studies.24,25,27 Unlike the paired-value difference approach outlined earlier, EG analysis focuses on evaluating the clinical risk associated with the distribution of results between the investigated system and the reference method. This evaluation involves categorizing pairs of glucose measurement results into risk scores (e.g. surveillance error grid [SEG]) or risk zones (e.g. consensus and continuous glucose GDs).34 Clark EG analysis for both rtCGM systems showed that 98% of the paired rtCGM-POC readings were within acceptable zones A and B, while for Medtronic Guardian Connect, 100% of the readings were within zones A and B.24,25 In contrast, the analysis of Dexcom G6 resulted in 0.2% of pairs in zone C (readings may have the potential to result in inappropriate treatment due to the clinical impact of hyperglycaemia or hypoglycaemia), 1.2% within zone D (undetected hyperglycaemia or hypoglycaemia that requires treatment) and 0.4% within zone E (hypoglycaemia mistaken for hyperglycaemia and vice versa), while in the study of Sadhu et al. only 2% of values were within zone D.24,25 An SEG analysis of Dexcom G6 showed that 82.9% of rtCGM-POC matched pairs were in zone 0 (no risk), 14.1% in zone 1 (slight, lower risk), 2.7% in zone 2 (slight, higher risk) and 0.4% in zone 3 (moderate, lower risk).27

Only one study reported the systematic bias, which illustrates the agreement between rtCGM-POC pairs, while Sadhu et al. used the Bland–Altman plot for their analysis.24,27 The systematic bias was 3.5% for rtCGM-POC pairs, 0.4% for rtCGM-Lab pairs and 2.7% for rtCGM-POC-Lab pairs. The 95% limits of agreement were -30.2 to +37.2% for rtCGM-POC, -25.2 to +26% for rtCGM-Lab and -29.3 to +34.7% for rtCGM-POC-Lab pairs.27 In the Bland–Altman analysis, Dexcom G6 showed a mean bias of -1.94 mg/dL (p<0.001) with 95% limits of agreement between -48.74 and +44.86 mg/dL. Medtronic Guardian Connect had a mean bias of -17.76 mg/dL with 95% limits of agreement between -72.91 and +37.4 mg/dL.24 In the study by Longo et al., rtCGM demonstrated better accuracy measures, such as MARD, SEG analysis, mean bias and 95% agreement, when compared with Lab values (p=0.0169) rather than POC values.27 The overall MARD for critically ill patients was 12.7% with POC reference pairs and 10.4% with glucose reference pairs.27 Two studies assessed the concordance between rtCGM and POC measurements.23,28 Davis et al. found that for sensor glucose values above 100 mg/dL, rtCGM measurements were within 20% of the corresponding POC reference values in about 75.7% of cases. They also noted that rtCGM glucose values were lower than POC measurements during hypoperfusion states, such as shock, pulseless electrical activity, therapeutic hypothermia protocols and conditions causing sensor compression, including pronation and bathing.23 Similar findings were observed during compression from blood pressure cuffs.24 Chow et al. reported high concordance between rtCGM and arterial POC values in 93% of critically ill patients, even with significant glycaemic variability and elevated glucose levels.28 None of the studies reviewed reported that mechanical ventilation, steroid therapy, vasopressors, muscle relaxants, albuterol, ascorbic acid, high doses of acetaminophen, hydroxyurea or renal replacement therapy contributed to inaccuracies in CGM readings.

Reduction of point-of-care measurements

Four studies assessed the reduction in POC measurements due to the use of rtCGM.23–26 On average, there was a 56.77% reduction in POC glucose testing, compared with each institution’s guideline frequency, while three studies demonstrated a reduction greater than 60%.23,25,26 Specifically, Sadhu et al. showed that using rtCGM systems led to a significant reduction in the frequency of POC testing. On the second day of using the rtCGM system for clinical management, the mean POC testing frequency decreased by 33.11% compared with day 0, when rtCGM values were not used. In this study, although there were POC testing frequency guidelines, there was inconsistency in nursing compliance with excessive POC testing.24 In another study, it was reported a reduction in POC testing in 50% of the participants, with a simultaneous increase in mean daily POC testing from 5.5 (prior to rtCGM use) to 6.9 ± 3.8 (during the use of rtCGM). This increase was attributed to the prolonged visit to the patient room for non-diabetes care, unfamiliarity and distrust of nurses in innovative monitoring technologies and the endless effort to understand COVID-19.28 In a study by Faulds et al., a 71% reduction in POC glucose testing was observed, with the biggest reduction occurring from the first day (mean POC testing=12) to the second day of rtCGM use (mean POC testing=7) and smaller reductions occurring daily until the fifth day (mean POC testing=4.9).26 Furthermore, in another study by Agarwal et al., the necessary POC testing was reduced by 60% during Dexcom G6 use, while similar findings of a 63% reduction in POC glucose testing were reported by Davis et al. when compared with the institution’s POC testing guidelines.23,25–28,34

Two studies evaluated the changes in mean glucose values during rtCGM for glucose management in hospitalized patients.26,28 Chow et al. assessed the differences in mean glucose levels measured by CGM both during (mean glucose 202.7 ± 42.1 mg/dL) and prior to the introduction of rtCGM (mean glucose 235.7 ± 42.1 mg/dL). Their results demonstrated a 14% reduction in mean glucose levels (p=0.0003) following the implementation of rtCGM, emphasizing the potential of this technology to improve glucose control in a hospital setting.28 In another study, Faulds et al. found that the average sensor glucose levels progressively decreased over time, with a 2.95% reduction observed when comparing glucose levels between the first day of monitoring and days 2–7.26 Specifically, the mean sensor glucose on the first day was 170 mg/dL, which decreased to 165 mg/dL between days 2 and 7.

Feasibility

The feasibility of rtCGM systems was evaluated in clinical trials by examining factors such as sensor placement ease, the reliability of remote data display, app usability and external influences on data collection.

Sadhu et al. found that both Dexcom G6 and Medtronic Guardian Connect required a lengthy initiation process, including account creation on a cloud-based platform. However, sensor insertion was quick (under 5 minutes), and the display setup took 10 minutes. The Medtronic system required additional procedures and daily calibrations, while the Dexcom G6 app allowed easier input of POC glucose values.24 Faulds et al. reported six cases of early sensor removal or reinsertion, including accidental sensor removal (n=1), placement error (n=1), concerns about sensor accuracy (n=1) and patient discharge or death (n=3).26 Some complications included minimal bleeding at the insertion site.24 Davis et al. reported that one patient experienced bleeding at the sensor placement site, which subsequently led to sensor failure, and defibrillation during cardiac arrest caused data transmission gaps.23

Reliability

The reliability of rtCGMs was evaluated by examining interruptions or gaps in the transmission of glucose data to the display device, or early sensor failure leading to sensor replacement. In a study by Sadhu et al., no data gaps or early sensor failure were reported in patients using Dexcom G6, while there was one loss of connection between the Dexcom G6 sensor and the display device and also two instances requiring sensor replacement.24 Furthermore, data gaps were documented during a computed tomography (CT) scan, and a sensor had to be removed for a magnetic resonance imaging (MRI) scan. In another study, after the initial placement and warm-up period, one Dexcom G6 sensor falsely displayed low measurements ranging from 40 to 55 mg/dL for 3.5 hours, resulting in early sensor replacement despite multiple calibrations.26

Time in, above and below range

Three studies assessed the percentage changes in time in range (TIR), time above range (TAR) and time below range (TBR), while TBR for glucose values less than 55 mg/dL was reported in one study.23,25,26,28 On average, patients had a 63.1% TIR. Additionally, two other studies concluded that the TIR was at least 71.4%.23,26 The mean TIR (70–180 mg/dL) of Dexcom G6 was assessed in a study by Davis et al. It was reported that the TIR (70–180 mg/dL) was 71.4 ± 13.9%, the TAR for glucose values 180–250 mg/dL was 19.8 ± 9.7%, the TAR for glucose values >250 mg/dL was 7.5 ± 7.3% and the TBR (<70 mg/dL) was 0.6 ± 0.9%.23 In another study by Faulds et al., the TIR increased from 64 ± 23% on the first day of rtCGM use to 72 ± 16% on the second to seventh days. Concurrently, the TAR (>180 mg/dL) decreased from 34 ± 24% to 28 ± 15%, and the TBR (<70 mg/dL) decreased from 1.5 ± 4.1% to 0.16 ± 0.35%. In the same study, none of the sensor glucose values below 70 mg/dL had a corresponding POC value below 70 mg/dL, and there was only one POC measurement within the hypoglycaemic range (<54 mg/dL).26 Agarwal et al. found a lower TIR (46.1 ± 15.8%), higher TAR (53.3 ± 15.8%) and similar TBR (0.6 ± 0.8%) compared with other studies.25 In the three studies that reported percentages, the TBR was consistently ≤0.6%. In the study by Chow et al., only three patients spent 1.7–2.6% of their time below 55 mg/dL, while the other participants spent less than 1% of their time within the hypoglycaemic range for glucose values <55 mg/dL.28

Nursing acceptance

In the study by Chow et al., 63% of the nurses involved in rtCGM use reported a significant improvement in the care of patients with COVID-19 and diabetes during the utilization of the devices, while 49% reduced their use of PPE.28 In another study, although a formal questionnaire was not used, nurses did not find rtCGM device insertion and maintenance to be a burden. Both rtCGM devices (Dexcom G6 and Medtronic Guardian Connect) were well accepted, and the ability to monitor hypo- and hyperglycaemic events was considered even better than frequent POC blood glucose testing.24 Additionally, integrating rtCGM systems into the ICU environment created a sense of safety when titrating insulin doses. A sense of safety was also observed in Faulds et al.’s study, where non-adjunctive insulin titrations, based on rtCGM values, increased from 41% of total insulin titrations on day 1 to 63% from the second to the fifth day of use.26 In many cases, there was an increased demand for more patients to use the devices. In the same study, nursing staff remained sceptical about the absolute differences between rtCGM and POC values and continued to perform POC testing according to the institution’s usual insulin infusion protocol.24,26,27

Discussion

Although there are no established guidelines for an acceptable MARD value in critically ill patients, studies suggest that a CGM MARD value <14% is generally acceptable, while a MARD >18% indicates poor accuracy.35,36 The mean MARD reported in this systematic review (12.47%) aligns with the MARD values observed in studies that used Dexcom G6 and Medtronic Guardian Connect in non-critically ill patients (9.0–12.8%).32,37–41 While accuracy has not yet reached the FDA approval standards required for the non-adjunctive use of these devices in the ICU setting, the results remain promising, particularly considering the severity of illness in the patient population and the low systematic bias.

Critical illness, including factors such as mechanical ventilation, vasopressors, renal replacement therapy, hypoxia, acidosis, fever, corticosteroids and enteral and parenteral nutrition, can induce high glycaemic variability. Despite studies including patients with high Sequential Organ Failure Assessment scores, refractory hyperglycaemia and increased insulin requirements, hybrid protocols using rtCGM resulted in acceptable glycaemic control with a low incidence of hypoglycaemia, as evidenced by the time in, above and below range data.

While CGM technology has evolved over the years and overcome several limitations, it has not yet reached the point where it can completely replace POC glucose testing. rtCGM devices are still not fully capable of detecting hypo- or hyperglycaemia in certain situations (zone D), which may lead to mistreatment (zone E). Clinical factors such as hypoperfusion, sensor compression, hypothermia protocols, acetaminophen, hydroxyurea and imaging tests (e.g. MRI) can induce discrepancies between POC and rtCGM measurements. As a result, most studies have shown a reduction of up to 50% in POC-testing frequency. This reduction not only alleviates healthcare providers’ exposure to SARS-CoV-2 but also reduces their workload.

Moreover, some of the studies included patients with medical conditions previously excluded from traditional protocols, such as those requiring mechanical ventilation, steroid therapy, vasopressors, muscle relaxants, albuterol, ascorbic acid, high doses of acetaminophen and hydroxyurea, renal replacement therapy, DKA and hyperosmolar hyperglycaemic coma. These conditions are important to be considered as they can affect glucose measurements and clinical outcomes.

Recent studies have highlighted the significant role of real-time continuous glucose monitoring in managing inpatient hyperglycaemia, particularly in critically ill patients. This technology has the potential to minimize the risks associated with traditional BGM, such as delayed results and inconsistent accuracy, while also improving patient outcomes by offering continuous data on glucose trends. As hospitals look to enhance glycaemic control in high-risk patients, the implementation of rtCGM is a promising advancement in inpatient care.42,43

Conclusions

A hybrid protocol that combines rtCGM with intermittent POC measurements appears to be a feasible and safe alternative for critically ill patients. This approach helps reduce healthcare providers’ exposure to SARS-CoV-2, preserves PPE, maintains optimal glycaemic control and enables real-time glucose monitoring. Additionally, it seems to reduce patient discomfort, blood loss and sleep interruptions. Continuous glucose monitoring can assist in identifying dangerous glycaemic changes and improve our understanding of the impact of COVID-19 on glycaemic variability. However, further studies, particularly randomized controlled trials, are needed to assess the safety, accuracy and cost-effectiveness of rtCGM in larger ICU populations. Additionally, modifications to rtCGM technology will be necessary for broader clinical use.

3

References

List View
Grid View
1

 Cascella MRajnik MAleem Aet alFeatures, evaluation, and treatment of coronavirus (COVID-19). In: StatPearls. Treasure Island, FL: StatPearls Publishing2022.

2
Copy DOIDOI Copied
Visit DOI Link

 Richardson SHirsch JSNarasimhan Met alPresenting characteristics, comorbidities, and outcomes among 5700 patients hospitalized with COVID-19 in the New York city area. JAMA. 2020;323:20529. DOI: 10.1001/jama.2020.6775.

3
Copy DOIDOI Copied
Visit DOI Link

 Kumar AArora ASharma Pet alIs diabetes mellitus associated with mortality and severity of COVID-19? A meta-analysisDiabetes Metab Syndr2020;14:53545. DOI: 10.1016/j.dsx.2020.04.044.

4
Copy DOIDOI Copied
Visit DOI Link

 Guo WLi MDong Yet alDiabetes is a risk factor for the progression and prognosis of COVID-19Diabetes Metab Res Rev. 2020;36:e3319DOI10.1002/dmrr.3319.

5
Copy DOIDOI Copied
Visit DOI Link

 Fadini GPMorieri MLLongato Eet alPrevalence and impact of diabetes among people infected with SARS-CoV-2J Endocrinol Invest2020;43:8679DOI10.1007/s40618-020-01236-2.

6
Copy DOIDOI Copied
Visit DOI Link

 Gregory JMSlaughter JCDuffus SHet alCOVID-19 severity is tripled in the diabetes community: A prospective analysis of the pandemic’s impact in type 1 and type 2 diabetes. Diabetes Care. 2021;44:52632. DOI10.2337/dc20-2260.

7
Copy DOIDOI Copied
Visit DOI Link

 Brufsky AHyperglycemia, hydroxychloroquine, and the COVID-19 pandemicJ Med Virol. 2020;92:7705. DOI: 10.1002/jmv.25887.

8
Copy DOIDOI Copied
Visit DOI Link

 Ceriello AZarich SWTesta RLowering glucose to prevent adverse cardiovascular outcomes in a critical care settingJ Am Coll Cardiol2009;53:S9S13. DOI: 10.1016/j.jacc.2008.09.054.

9
Copy DOIDOI Copied
Visit DOI Link

 Klonoff DCMessler JCUmpierrez GEet alAssociation between achieving inpatient glycemic control and clinical outcomes in hospitalized patients with COVID-19: A multicenter, retrospective hospital-based analysis. Diabetes Care. 2021;44:57885. DOI: 10.2337/dc20-1857.

10
Copy DOIDOI Copied
Visit DOI Link

 Yao RQRen CWu GSet alIs intensive glucose control bad for critically ill patients? A systematic review and meta-analysis. Int J Biol Sci2020;16:165875. DOI: 10.7150/ijbs.43447.

11
Copy DOIDOI Copied
Visit DOI Link

 Peric SStulnig TMDiabetes and COVID-19: Disease-management-peopleWien Klin Wochenschr2020;132:35661. DOI: 10.1007/s00508-020-01672-3.

12
Copy DOIDOI Copied
Visit DOI Link

 Kim NYHa EMoon JSet alAcute hyperglycemic crises with coronavirus disease-19: Case reportsDiabetes Metab J. 2020;44:34953. DOI10.4093/dmj.2020.0091.

13
Copy DOIDOI Copied
Visit DOI Link

 Umpierrez GEHellman RKorytkowski MTet alManagement of hyperglycemia in hospitalized patients in non-critical care setting: An endocrine society clinical practice guidelineJ Clin Endocrinol Metab2012;97:1638. DOI10.1210/jc.2011-2098.

14
Copy DOIDOI Copied
Visit DOI Link

 Clement SBraithwaite SSMagee MFet alManagement of diabetes and hyperglycemia in hospitalsDiabetes Care. 2004;27:55391. DOI: 10.2337/diacare.27.2.553.

15
Copy DOIDOI Copied
Visit DOI Link

 Cobaugh DJMaynard GCooper Let alEnhancing insulin-use safety in hospitals: Practical recommendations from an ASHP foundation expert consensus panelAm J Health Syst Pharm. 2013;70:140413. DOI: 10.2146/ajhp130169.

16
Copy DOIDOI Copied
Visit DOI Link

 Le HTHarris NSEstilong AJet alBlood glucose measurement in the intensive care unit: What is the best method? J Diabetes Sci Technol2013;7:48999. DOI: 10.1177/193229681300700226.

17
Copy DOIDOI Copied
Visit DOI Link

 Cheung NWNapier BZaccaria Cet alHyperglycemia is associated with adverse outcomes in patients receiving total parenteral nutritionDiabetes Care2005;28:236771. DOI: 10.2337/diacare.28.10.2367.

18
Copy DOIDOI Copied
Visit DOI Link

 Adolfsson PParkin CGThomas Aet alSelecting the appropriate continuous glucose monitoring system – A practical approachEur Endocrinol2018;14:249. DOI: 10.17925/EE.2018.14.1.24.

19
Copy DOIDOI Copied
Visit DOI Link

 Gianchandani REsfandiari NHAng Let alManaging hyperglycemia in the COVID-19 inflammatory storm. Diabetes. 2020;69:204853. DOI: 10.2337/dbi20-0022.

20
Copy DOIDOI Copied
Visit DOI Link

 Pasquel FJUmpierrez GEIndividualizing inpatient diabetes management during the coronavirus disease 2019 pandemicJ Diabetes Sci Technol2020;14:7057. DOI: 10.1177/1932296820923045.

21
Copy DOIDOI Copied
Visit DOI Link

 DexcomFact Sheet for Healthcare Providers: Use of Dexcom Continuous Glucose Monitoring Systems in the HospitalAvailable at: www.dexcom.com/hospitalfacts (accessed: 10 July 2024).

22
Copy DOIDOI Copied
Visit DOI Link

 US Department of Health and Human Services, Food and Drug Administration, Center for Devices and Radiological HealthEnforcement Policy for Non-Invasive Remote Monitoring Devices Used to Support Patient Monitoring During the Coronavirus Disease 2019 (COVID-19) Public Health Emergency: Guidance for Industry and Food and Drug Administration Staff (March 2020), FDA-20202-D-1138. Available at: www.fda.gov/regulatory-information/search-fda-guidance-documents/enforcement-policy-non-invasive-remote-monitoring-devices-used-support-patient-monitoring (accessed: 10 July 2024).

23
Copy DOIDOI Copied
Visit DOI Link

 Davis GMFaulds EWalker Tet alRemote continuous glucose monitoring with a computerized insulin infusion protocol for critically ill patients in a COVID-19 medical ICU: Proof of conceptDiabetes Care2021;44:10558. DOI: 10.2337/dc20-2085.

24
Copy DOIDOI Copied
Visit DOI Link

 Sadhu ARSerrano IAXu Jet alContinuous glucose monitoring in critically ill patients with COVID-19: Results of an emergent pilot studyJ Diabetes Sci Technol. 2020;14:106573. DOI: 10.1177/1932296820964264.

25
Copy DOIDOI Copied
Visit DOI Link

 Agarwal SMathew JDavis GMet alContinuous glucose monitoring in the intensive care unit during the COVID-19 pandemicDiabetes Care2021;44:8479. DOI: 10.2337/dc20-2219.

26
Copy DOIDOI Copied
Visit DOI Link

 Faulds ERBoutsicaris ASumner Let alUse of continuous glucose monitor in critically ill COVID-19 patients requiring insulin infusion: An observational studyJ Clin Endocrinol Metab. 2021;106:e400716. DOI: 10.1210/clinem/dgab409.

27
Copy DOIDOI Copied
Visit DOI Link

 Longo RRElias HKhan Met alUse and accuracy of inpatient CGM during the COVID-19 pandemic: An observational study of general medicine and ICU patientsJ Diabetes Sci Technol. 2022;16:113643. DOI: 10.1177/19322968211008446.

28
Copy DOIDOI Copied
Visit DOI Link

 Chow KWKelly DJRieff MCet alOutcomes and healthcare provider perceptions of real-time continuous glucose monitoring (rtCGM) in patients with diabetes and COVID-19 admitted to the ICUJ Diabetes Sci Technol. 2021;15:60714. DOI: 10.1177/1932296820985263.

29
Copy DOIDOI Copied
Visit DOI Link

 Beck RWRiddlesworth TRuedy Ket alEffect of continuous glucose monitoring on glycemic control in adults with type 1 diabetes using insulin injections: The DIAMOND randomized clinical trialJAMA2017;317:3718. DOI: 10.1001/jama.2016.19975.

30
Copy DOIDOI Copied
Visit DOI Link

 Beck RWRiddlesworth TDRuedy Ket alContinuous glucose monitoring versus usual care in patients with type 2 diabetes receiving multiple daily insulin injections: A randomized trial. Ann Intern Med. 2017;167:36574. DOI: 10.7326/M16-2855.

31
Copy DOIDOI Copied
Visit DOI Link

 Charleer SMathieu CNobels Fet alEffect of continuous glucose monitoring on glycemic control, acute admissions, and quality of life: A real-world studyClin Endocrinol Metab. 2018;103:122432. DOI: 10.1210/jc.2017-02498.

32
Copy DOIDOI Copied
Visit DOI Link

 Reutrakul SGenco MSalinas Het alFeasibility of inpatient continuous glucose monitoring during the COVID-19 pandemic: Early experienceDiabetes Care2020;43:e1378. DOI: 10.2337/dc20-1503.

33
Copy DOIDOI Copied
Visit DOI Link

 Galindo RJAleppo GKlonoff DCet alImplementation of continuous glucose monitoring in the hospital: Emergent considerations for remote glucose monitoring during the COVID-19 pandemicJ Diabetes Sci Technol. 2020;14:82232. DOI: 10.1177/1932296820932903.

34
Copy DOIDOI Copied
Visit DOI Link

 Freckmann GPleus SGrady Met alMeasures of accuracy for continuous glucose monitoring and blood glucose monitoring devicesJ Diabetes Sci Technol2019;13:57583. DOI10.1177/1932296818812062.

35
Copy DOIDOI Copied
Visit DOI Link

 Wollersheim TEngelhardt LJPachulla Jet alAccuracy, reliability, feasibility and nurse acceptance of a subcutaneous continuous glucose management system in critically ill patients: A prospective clinical trialAnn Intensive Care. 2016;6:70. DOI: 10.1186/s13613-016-0167-z.

36
Copy DOIDOI Copied
Visit DOI Link

 Krinsley JSChase JGGunst Jet alContinuous glucose monitoring in the ICU: Clinical considerations and consensus. Crit Care. 2017;21:197. DOI: 10.1186/s13054-017-1784-0.

37
Copy DOIDOI Copied
Visit DOI Link

 Bailey TSChang AChristiansen MClinical accuracy of a continuous glucose monitoring system with an advanced algorithmJ Diabetes Sci Technol2015;9:20914. DOI: 10.1177/1932296814559746.

38
Copy DOIDOI Copied
Visit DOI Link

 Welsh JBGao PDerdzinski Met alAccuracy, utilization, and effectiveness comparisons of different continuous glucose monitoring systemsDiabetes Technol Ther. 2019;21:12832. DOI10.1089/dia.2018.0374.

39
Copy DOIDOI Copied
Visit DOI Link

 Christiansen MPGarg SKBrazg Ret alAccuracy of a fourth-generation subcutaneous continuous glucose sensorDiabetes Technol Ther2017;19:44656. DOI: 10.1089/dia.2017.0087.

40
Copy DOIDOI Copied
Visit DOI Link

 Nair BGDellinger EPFlum DRet alA pilot study of the feasibility and accuracy of inpatient continuous glucose monitoringDiabetes Care2020;43:e1689. DOI: 10.2337/dc20-0670.

41
Copy DOIDOI Copied
Visit DOI Link

 Tripyla AHerzig DJoachim Det alPerformance of a factory-calibrated, real-time continuous glucose monitoring system during elective abdominal surgeryDiabetes Obes Metab. 2020;22:167882. DOI: 10.1111/dom.14073.

42
Copy DOIDOI Copied
Visit DOI Link

 Guerrero-Arroyo LFaulds EPerez-Guzman MCet alContinuous glucose monitoring in the intensive care unitJ Diabetes Sci Technol2023;17:66778. DOI: 10.1177/19322968231169522.

43
Copy DOIDOI Copied
Visit DOI Link

 Patham BKansara ASadhu ARReal-time continuous glucose monitoring in the intensive care unit: The fifth vital signDiabetes Care2024;47:9246DOI10.2337/dci24-0020.

4

Article Information

Disclosure

John Doupis has received honoraria for lectures and consultation fees from Abbott. Georgios Logothetis, Konstantinos Avramidis, Evanthia Konstantaki and Vasiliki Matziou have no financial or non-financial relationships or activities to declare in relation to this article.

Compliance With Ethics

This article involves a review of the literature and did not involve any studies with human or animal subjects performed by any of the authors.

Review Process

Double-blind peer review.

Authorship

All named authors meet the criteria of the International Committee of Medical Journal Editors for authorship for this manuscript, take responsibility for the integrity of the work as a whole and have given final approval for the version to be published.

Correspondence

John DoupisInternal Medicine and Diabetes ClinicIatriko Palaiou Falirou Medical Center, 36 Areos Street, 17562, Paleo FaliroAthensGreecejohn.doupis@harvard-alumni.org

Support

No funding was received in the publication of this article.

Access

This article is freely accessible at touchENDOCRINOLOGY.com. ©Touch Medical Media 2025.

Data Availability

Data sharing is not applicable to this article as no datasets were generated or analysed during the current study/during the writing of this article.